Model Number 8780
Device Problem Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2021
Event Type  malfunction  
Manufacturer Narrative
Analysis identified an anomaly to the catheter/guide wire.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a health care professional (hcp) via a device manufacturer representative regarding a patient receiving saline via an implantable infusion pump.It was reported that during initial catheter placement, the hcp was unable to withdraw the stylet from the catheter.Multiple attempts were made, but the stylet could not be removed.A new catheter set was utilized and the issue was resolved.There were no environmental, external or patient factors which may have led or contributed to the issue.No further complications were reported.2021-07-07 rp: document contained no new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCENDA
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12438396
MDR Text Key270278465
Report Number3004209178-2021-13469
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00763000051112
UDI-Public00763000051112
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/19/2023
Device Model Number8780
Device Catalogue Number8780
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
Patient Weight122
-
-