| Model Number |
96-0088 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a rev knee done in 2017 by the same dr at the same hospital for the same problem.The femoral adaptor bolt and adaptor was broken leaving the femoral component decoupled from the sleeve stem construct.The femoral component, broken bolt and tibial insert were removed.We turned our attention to the adaptor that was fixed to the sleeve, after manuy attempts to break the taper between the adaptor and the sleeve we were unsuccessful.The stem, sleeve and remain piece of the adaptor bolt were removed and a distal femoral replacement was preformed.Femoral sleeve and stem were from original surgery in 2017, i do not have sizes or lot numbers for these implants.Doi: (b)(6) 2017 , dor: (b)(6) 2021, (right knee).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary - no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot - the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|