| Model Number |
N/A |
| Medical Device Problem Code |
Device Damaged Prior to Receipt by User (1570)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source: (b)(6).Complaint sample was evaluated and the reported event was confirmed.Sterility has been breached for lot 6209552, as the sterile pouch and blister have both been damaged.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event can be attributed to transit damage.
|
| |
|
Event or Problem Description
|
|
It was reported that damaged sterile packaging was identified in the warehouse.No further information is available.
|
| |
|