|
The returned product was subjected to a technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Only the distal portion of the delivery system and two stent fragments were returned for analysis.The 214 mm long device fragment shows a strong kink about 105 mm proximal to its distal end.The outer shaft is retracted by about 44 mm.Outside of the deformed zone the outer shaft diameter still complies with the specification.One stent fragment was located under the retractable outer shaft whereas the other stent fragment was located inside the distal portion of an introducer sheath fragment.Both stent fragments are severely deformed, indicating application of a high tensile force.Due to the state of the instrument, an investigation with regards to the reported complaint event is not possible.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The final root cause could not be established.It should be noted that, according to the ifu, the pulsar-18 stent system is indicated for use in the superficial femoral or proximal popliteal arteries.
|