| Model Number |
N/A |
| Medical Device Problem Codes |
Overheating of Device (1437); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the cart would not power on when plugged into the wall; no continuity through power inlet module.The power inlet module was replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that the cart was not powering on via the wall.The event timing was during cleaning.There was no harm to the patient.Investigation found there was burning/charring of a resistor in the fuse box; caused by a loose/poor fuse connection generating heat.No adverse event was reported as a result of this malfunction.
|
| |
|