MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPEED 15X12X12MM IMPLANT; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPEED 15X12X12MM IMPLANT; STAPLE, FIXATION, BONE Back to Search Results
Model Number SE-1512
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that, during a procedure, speed compression implant kit deployed from inserter before product was ready to be implanted.Surgical staff attempted to reattach the implant to the inserter and the implant fell off on the floor.Concomitant devices reported: unknown insertion instrument (part# unknown, lot# unknown, quantity 1).This report is for one (1) speed 15x12x12mm implant.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: part: se-1512.Synthes lot: bse200031.Supplier lot: n/a.Release to warehouse date: february 13, 2020.Expiration date: january 27, 2025.Manufactured by: bme.No nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEED 15X12X12MM IMPLANT
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12440500
Report Number8030965-2021-07566
Device Sequence Number3613055
Product Code JDR
UDI-Device Identifier00810633020166
UDI-Public(01)00810633020166
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K142292
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2025
Device Model NumberSE-1512
Device Catalogue NumberSE-1512
Device Lot NumberBSE200031
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/10/2021
Supplement Date Received by Manufacturer10/01/2021
Initial Report FDA Received Date09/08/2021
Supplement Report FDA Received Date10/05/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - INSERTION INSTRUMENT; UNK - INSERTION INSTRUMENT
Patient Age41 YR
-
-