| Model Number |
H7493941838350 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: a synergy xd mr us 3.50 x 38mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage.Stent struts on the distal end were lifted and pulled distally.The undamaged stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 18aug2021.It was reported that crossing difficulties were encountered.A 3.50 x 38mm synergy xd drug-eluting stent was advanced but failed to cross the lesion.No patient complications were reported.However, device analysis revealed stent damage.
|
| |
|
Search Alerts/Recalls
|