• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OSTEORAPTOR 2.3 W/ 1 UB COBRAID BLACK; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OSTEORAPTOR 2.3 W/ 1 UB COBRAID BLACK; SCREW, FIXATION, BONE Back to Search Results
Model Number 72201992
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, during an unknown procedure, the thread of the osteoraptor 2.3 suture anchor got cut near the anchor during tying the knot.The procedure was successfully completed without significant delay using a back-up device.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A complaint history review concluded this was an isolated event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A review of the suture specifications found that a material certification is required with each lot.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEORAPTOR 2.3 W/ 1 UB COBRAID BLACK
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key12442065
Report Number1219602-2021-01892
Device Sequence Number14307932
Product Code HWC
UDI-Device Identifier03596010615534
UDI-Public03596010615534
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/16/2025
Device Model Number72201992
Device Catalogue Number72201992
Device Lot Number2061385
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/18/2021
Supplement Date Received by Manufacturer10/12/2021
Initial Report FDA Received Date09/08/2021
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-