Event or Problem Description
It was reported that stent dislodgement occurred.The 80% stenosed target lesion was located in the mildly tortuous and severely calcified vessel.A 3.50 x 24mm synergy xd drug-eluting stent was advanced for treatment.However, when the stent was being pulled back into a non-boston scientific (bsc) guide extension catheter, the stent came off balloon and was left in the proximal portion of right coronary artery.Multiple balloons were inserted through the stent starting with a 1.5mm and increasing in size.The physician then deployed a non-bsc stent to crush the dislodged stent.The procedure was completed successfully.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12442680
Report Number2134265-2021-11003
Device Sequence Number11975273
Product Code NIQ
UDI-Device Identifier08714729981060
UDI-Public08714729981060
Combination Product (Y/N)Y
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2021
Device Model NumberH7493941824350
Device Lot Number0027061917
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/23/2021
Initial Report FDA Received Date09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-