| Model Number |
3256 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the console screen flickered then faded to black.A rotapro console was selected for use.During start up and during the procedure, the console screen flickered and faded to black.The procedure was completed with a 1.5 burr.The patient's status was okay post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the console was returned in over all good physical condition.The console was powered up.After 15 minutes, the screen started to flicker.After 1 plus hours, the screen went blank resulting in a white display.The printed circuit assembly (pca) main board had failed.The pca is part of the front housing assembly.
|
| |
|
Event or Problem Description
|
|
It was reported that the console screen flickered then faded to black.A rotapro console was selected for use.During start up and during the procedure, the console screen flickered and faded to black.The procedure was completed with a 1.5 burr.The patient's status was okay post procedure.
|
| |
|