PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0101
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-01981, 3005168196-2021-01983.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the right middle cerebral artery (mca) bifurcation using penumbra smart coils (smart coil), a penumbra smart coil detachment handle (handle), a non-penumbra microcatheter, and a guidewire.It was noted that the aneurysm was bi-lobed.During the procedure, the physician successfully implanted two smart coils in the larger lobe of the aneurysm using the handle.The physician then advanced another smart coil into the target location and, while positioning the smart coil in the target location, the physician experienced resistance and kinked the pusher assembly of the smart coil.Therefore, the smart coil was retracted and removed.Subsequently, the physician decided to stop coiling the larger lobe of the aneurysm.Next, the physician successfully implanted a smart coil in the smaller lobe of the aneurysm using the handle.While attempting to detach another smart coil in the target location using the handle, the smart coil failed to detach.Subsequently, the physician attempted to reposition the smart coil.However, while repositioning, the physician lost access with the microcatheter.Therefore, the smart coil was also retracted and removed.The physician was satisfied with the results, and the procedure was completed at this point.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12443630
Report Number3005168196-2021-01982
Device Sequence Number11984912
Product Code HCG
UDI-Device Identifier00814548015590
UDI-Public00814548015590
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/08/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2025
Device Model Number400SMTHXSFT0101
Device Catalogue Number400SMTHXSFT0101
Device Lot NumberF98547
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2021
Initial Report FDA Received Date09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Date Report Sent to FDA01/01/2005
-
-