Model Number DIU150
Medical Device Problem Codes Crack (1135); Failure to Advance (2524); Mechanical Jam (2983); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient weight, ethnicity and race: unknown as information was not provided.Date implanted: not applicable as the iol was not implanted.Date explanted: not applicable as the iol was not implanted.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported the pre-loaded intraocular lens (iol) had lens scratch, loading, stuck in cartridge and cartridge tip cracked/deformed issues.The cartridge tip had contact with the patient's ocular dexter (right eye).The procedure was completed using another lens with the same model and diopter size.There was no medical intervention and no surgical intervention.Outcome does not significantly interfere with activities of daily life.Patient status post-procedure is unknown.No further information is available.
 
Additional Manufacturer Narrative
Additional information: section h-3: device evaluated by manufacturer: yes.Device evaluation: the original folding carton, patient stickers, and additional information was received on the patient stickers.However, complaint product was not received, and therefore no product evaluation could be performed on the lens.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.There are no discrepancies found during the mrr (manufacturing record review) related to this complaint.Historical data analysis: a search revealed that no other complaints have been received for this production order (po) number.Conclusion: as a result of the investigation, there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
1700 e. st. andrew place
santa ana, CA 92705
7142478552
MDR Report Key12444839
Report Number2020664-2021-07165
Device Sequence Number3594045
Product Code MJP
UDI-Device Identifier05050474746572
UDI-Public(01)05050474746572(17)240206
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/06/2024
Device Model NumberDIU150
Device Catalogue NumberDIU150U215
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2021
Supplement Date Received by Manufacturer10/24/2021
Initial Report FDA Received Date09/09/2021
Supplement Report FDA Received Date11/17/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-