Model Number 5703690
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 03/2023).
 
Event or Problem Description
It was reported that during a procedure, the catheter was allegedly broken.It was further reported that blood leak occurred.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a procedure, the catheter was allegedly broken.It was further reported that blood leak occurred.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation, however one photo was provided for review.The investigation is confirmed for the reported fluid leak issue as blood leakage was found on the catheter just distal to the bifurcation.However, the investigation is inconclusive for the reported fracture as the provided photo itself is not clear enough to confirm the issue.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 03/2023).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPLIT LONG-TERM HEMODIALYSIS CATHETER, POLYURETHANE, KIT, 14.5F
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key12444951
Report Number3006260740-2021-03724
Device Sequence Number17039505
Product Code MSD
UDI-Device Identifier00801741012884
UDI-Public(01)00801741012884
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K030020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5703690
Device Catalogue Number5703690
Device Lot NumberREEW1673
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/13/2021
Supplement Date Received by Manufacturer09/15/2021
Initial Report FDA Received Date09/09/2021
Supplement Report FDA Received Date09/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-