• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BONE ANCHORS 3 W ARTHRO DEL SYSTEM; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BONE ANCHORS 3 W ARTHRO DEL SYSTEM; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 4403
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Loss of Range of Motion (2032); Skin Inflammation/ Irritation (4545)
Date of Event 03/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that after procedure, where regeneten bioinductive implant was used, the patient had pain and irritation on (b)(6) 2021.On a second surgery, it was noticed that the device become dislodged and was not secured to the supraspinatus tendon.It had bunched up and was partially attached to the acromion.The implant , bone anchors and tendon anchors were removed.The patient outcome is unknown.
 
Event or Problem Description
It was reported that on (b)(6) 2021 patient had a left shoulder arthroscopic rotator cuff repair using a regeneten patch.After this the patient presented pain and had to be treated with medication, injections and phyisical therapy.On (b)(6) 2021 the patient had a revision surgery where there was evidence of a midsubstance rupture of the bio inductive patch with tissue flipped on itself in the lateral glutter and also a ridge on the acromion which may have corresponded to the contact of the graft.The rupture of the patch was debrided with a combination of a shaver and grasper from the lateral portal, the 2 bone staples were removed and the soft tissue staples were also removed with a combination of a grasper and a shaver.The ridge was smoothed with a shaver.After this procedure patient presented pain again, medications were also provided.Current status of the patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference case-2021-00069452-3.D4: part and udi number updated.
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that based on the information provided, the definitive clinical root cause of the reported adverse event could not be determined.However, we cannot rule out the patient¿s non-compliance with the rehabilitation plan as a potential contributory factor to the reported events.Approximately seven-months post procedure this patient reports she is doing her home exercises and feels like she is improving.The patient rates her pain as 2 on a scale of 10, and she is happy with her progress.Therefore, the patient impact beyond what was reported could not be determined.All documents and images provided as of this date have been reviewed and considered, no further clinical/medical assessment is warranted at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Correction in h6 (health effect - clinical and impact codes).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE ANCHORS 3 W ARTHRO DEL SYSTEM
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12445088
Report Number3003604053-2021-00244
Device Sequence Number19712097
Product Code MBI
UDI-Device Identifier00854501006173
UDI-Public00854501006173
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K131635
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2023
Device Model Number4403
Device Catalogue Number2503-A
Device Lot Number2065152
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2021
Supplement Date Received by Manufacturer01/24/2022
11/04/2022
06/01/2023
Initial Report FDA Received Date09/09/2021
Supplement Report FDA Received Date02/06/2022
11/11/2022
06/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-