A physician reported a certas valve was implanted in the patient via l-p shunt on an unknown date with unknown setting.The valve was used with the silascon lumbar catheter (manufactured by kaneka, product code: 702-jj).The valve was reversed, so revision surgery was performed.The valve was removed and replaced on (b)(6) 2021.
|
Updated fields: d4, d9, g3, g6, h2, h3, h6, h10.The certas valve was returned for evaluation.Failure analysis - the position of the cam when valve was received was at setting 4.The valve was visually inspected; the needle guard was raised, also needle holes were noted in the underside of the needle chamber.The valve was hydrated.The valve was leak tested and leaked from the needle holes in the underside of the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.The needle chamber was dismantled; the needle guard was bent, and glue traces were noted on the needle guard and the silicone base.The root cause for the issue reported by the customer is probably due to user¿s error during implantation.The root cause for the needle holes in the underside of the needle chamber is probably due to the valve turning over during/after implantation.The root cause for the raised needle guard is probably due to wrong handling as noted in the ifu: do not fold or bend the valve, folding or bending might cause rupture of the silicone housing, needle guard disc dislodgement or occlusion of the fluid pathway, at the time of the investigation no occlusion was noted.
|