Siemens has completed an investigation of the reported event.Assessment does not indicate a system failure or malfunction and no non-conformity was identified.Unfortunately, despite all efforts to this event, there is only limited information available regarding what occurred on site.The customer's medwatch is unfortunately very contradictory.Among other things, it talks about a supposed exchanged artis zeego interventional fluoroscopic x-ray machine that was returned for examination.The fact alone that the system is currently in operation and would be much too large and heavy for a possible return proves that this information cannot be correct and that there has been a misunderstanding.We have not received any parts for examination.This suggests that perhaps the syngo x workplace was meant, which is a standalone computer with software and which works together with the artis zeego.However, the investigation could not find any error in the syngo x workplace and it works as specified.Although the customer had reported that there had been problems sending 3d images to the pacs archiving system, everything was fine after a reboot.The error mentioned in the complaint has not been reported again.The occurrence rate of the mentioned error pattern has been checked.Any accumulation of errors or even a systematic error that would lead to corrective action of the installed base could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
|