Model Number 168-450
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformance noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Healthcare professional reported left side deflation.Device was explanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: crease fold, opening on posterior and white particles inner the shell.Leak test and microscopic analysis was performed which identified: sharp opening on posterior and non-penetrating nicks.A dimension measurement in the shell was performed which identify the thickness within specification.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a sharp opening on posterior assessed as an unidentified (tear) opening.
 
Event or Problem Description
Healthcare professional reported left side deflation.Device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 168 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
MDR Report Key12446527
Report Number9617229-2021-49875
Device Sequence Number14331848
Product Code FWM
UDI-Device Identifier10888628000865
UDI-Public10888628000865
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/07/2021
Device Model Number168-450
Device Catalogue Number168-450
Device Lot Number3061141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Initial Date Received by Manufacturer 08/10/2021
Supplement Date Received by Manufacturer10/04/2021
Initial Report FDA Received Date09/09/2021
Supplement Report FDA Received Date10/26/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/07/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-