| Model Number |
168-450 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: crease fold, opening on posterior and white particles inner the shell.Leak test and microscopic analysis was performed which identified: sharp opening on posterior and non-penetrating nicks.A dimension measurement in the shell was performed which identify the thickness within specification.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a sharp opening on posterior assessed as an unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device was explanted.
|
| |
|