| Model Number |
MI1000 MED-EL CONCERT PIN |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
Reportedly, the user no longer had hearing benefit with the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
Reportedly, the user no longer had hearing benefit with the device.The user has been re-implanted.
|
| |
|
Event or Problem Description
|
|
Reportedly, the user no longer had hearing benefit with the device.The user has been re-implanted on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The active electrode also showed some additional damage to the electrode wires, as being caused by minute device mobility.An accident or impact might have weakened the fixation of the implant, consequently leading to electrode mobility.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|