ST. JUDE MEDICAL, INC. INQUIRY STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
| Model Number |
IBI-81104 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a supraventricular tachycardia procedure, the catheter did not keep its curve and a delay occurred.A new catheter was opened and manipulated, however the physician preferred the feel of the first catheter.Therefore, the first catheter inserted again and was not able to advance into the coronary sinus.A livewire catheter was then used to complete the procedure with no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One decapolar, inquiry steerable diagnostic catheter was received for evaluation.The catheter was found to have the handle cap partially unscrewed.Once tightened, the catheter deflected when actuating the steering mechanism, and deflected in the correct shape according to specifications.The catheter was then able to maintain its curve upon release of the steering mechanism.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported deflection issue and subsequent delay was due to the handle cap being partially unscrewed.The cause for the position of the handle cap remains unknown.
|
| |
|