| Model Number |
NX029Z |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an issue with nx029z - as vega ps femoral comp.Cemented f4n r.According to the complaint description, during explant of the primary knee components, the femur came right off and surgeon noticed that there was not much bone cement adhered to the femoral component.Only a couple of small patches of cement could be seen on the posterior condyles.No cement visible on any other part of component.All bone cement was only adhered to the bone on the femur.A revision surgery was necessary.Additional information was not provided.Additional patient information is not available.The adverse event is filed under aag reference (b)(4).Involved components: nx125 - vega ps gliding surface t2/2+ 20mm - lot unknown , nx053z - as vega ps tibial plateau cemented t2 - lot 52542951.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.It could be possible that there were problems with the cement technique and due to this no bone cement was adhered to the femoral component.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 2(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|
| |
|
Search Alerts/Recalls
|