| Model Number |
C650-129A |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device has not yet been received for evaluation.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
| |
|
Event or Problem Description
|
|
As reported, during a bladder biopsy and fulguration diagnostic procedure, the acmi monopolar cords (c650-129a) used with a bugbee electrode (acmi 257f8) caught fire during use.The incident occurred after brief intermittent use of instruments for several minutes.The cord was replaced with a different cord of the same kind.The intended procedure was completed with no harm or injury to the patient.No user injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Upon inspection, the complaint was confirmed.Based on the results of the investigation, the reported failure is likely due to the device being used beyond service life.This issue is addressed in the instructions for use (ifu): "when used as intended, this product is subject to wear or damage depending on the intensity of use.Inspect the product prior to each use to ensure it has no externally visible defects, including wear, kinks, cuts or pinching.Using this product when damaged may be hazardous;" and "this product has a restricted service life.Do not use the cable for a period of more than one year.Using the cable beyond one year may be hazardous." (cautions, page 1).Olympus will continue to monitor the field performance of this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to correct information previously reported.Field d4 / lot number has been corrected to identify lot number "bp" for the subject device of this report.
|
| |
|