FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problems Break (1069); Crack (1135); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in (b)(6), via a fisher & paykel healthcare (f&p) field representative, that the collar of a rt380 adult dual heated evaqua2 breathing circuit was damaged during use.There was no reported patient consequence.
 
Event Description
A distributor reported on behalf of a healthcare facility in japan, via a fisher & paykel healthcare (f&p) field representative, that the collar of a rt380 adult dual heated evaqua2 breathing circuit was damaged during use.There was no reported patient consequence.
 
Manufacturer Narrative
Ps377223, method: the complaint rt380 adult dual heated evaqua2 breathing circuit was received at fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: visual inspection of the returned rt380 adult dual heated evaqua2 breathing circuit confirmed that the patient end evaqua2 collar was cracked.It was also noted that a white residue was found inside the y-piece.Further analysis of the returned complaint device revealed that the fractured surface of the collar had tree-ring like deformation pattern that is typical of environmental stress cracking which form due to chemical exposure or contact.Conclusion: the observed crack is most likely due to the collar being in contact with a chemical resulting in environmental stress cracking.All rt380 adult dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject circuits would have met the required specifications at the time of production.The user instructions that accompany the rt380 adult dual heated evaqua2 breathing circuit states: "do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "set appropriate ventilator alarms.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key12451638
MDR Text Key273191054
Report Number9611451-2021-01009
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
HUMIDIFIER: MR850GJU; HUMIDIFIER: MR850GJU; SAVINA VENTILATOR; SAVINA VENTILATOR
-
-