MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-45
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device : an authorized getinge dealer engineer was dispatched to investigate.The engineer evaluated the iabp unit, but was unable to reproduce the reported issue.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer, however, not cleared for clinical service.Repairs are ongoing, and a supplemental report will be submitted upon completion of our investigation.The full event site name in is (b)(6).
 
Event Description
It was reported that during use, the cardiosave intra-aortic balloon pump (iabp) alarmed "system temperature is too high," and continued to alarm even after the end user re-started the unit.No patient harm, serious injury or adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
An authorized getinge dealer engineer was dispatched to investigate.The engineer evaluated the iabp unit, but was unable to reproduce the reported issue.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12452427
MDR Text Key270866342
Report Number2249723-2021-02041
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-45
Device Catalogue Number0998-00-0800-45
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/18/2021
Initial Date FDA Received09/10/2021
Supplement Dates Manufacturer Received01/16/2024
Supplement Dates FDA Received01/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
Patient SexPrefer Not To Disclose
-
-