DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-45 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device : an authorized getinge dealer engineer was dispatched to investigate.The engineer evaluated the iabp unit, but was unable to reproduce the reported issue.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer, however, not cleared for clinical service.Repairs are ongoing, and a supplemental report will be submitted upon completion of our investigation.The full event site name in is (b)(6).
|
|
Event Description
|
It was reported that during use, the cardiosave intra-aortic balloon pump (iabp) alarmed "system temperature is too high," and continued to alarm even after the end user re-started the unit.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
An authorized getinge dealer engineer was dispatched to investigate.The engineer evaluated the iabp unit, but was unable to reproduce the reported issue.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|