Catalog Number UNKNOWN
Device Problems Difficult to Remove (1528); Structural Problem (2506)
Patient Problems Perforation of Vessels (2135); Anxiety (2328); Depression (2361)
Event Type  Injury  
Manufacturer Narrative
Occupation: non-healthcare professional.Investigation: the reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Catalog number and lot number are unknown.The alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
The following information is alleged: the patient received a gunther tulip inferior vena cava (ivc) filter on (b)(6) 2007.The patient's filter perforated through the inferior vena cava.Hospital and medical records have been requested, but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2007 due to deep vein thrombosis (dvt), followed by an unsuccessful retrieval attempt on (b)(6) 2008.Patient is alleging vena cava perforation.The patient further alleges anxiety, depression, bipolar disorder, worry, fear, and physical limitations.Report from ct (computed tomography): "this dictation refers only to the ivc filter.Cook gunther tulip filter.Caval perforation: yes.2 o'clock 7.10 mm grade 2.Tilt: no.Migration: no.Pertinent negatives: none.".
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: unable to remove, anxiety, depression, bipolar, worry, fear, physical limitations.Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up.The risks/benefits of filter retrieval should be considered for each patient during follow-up.Once protection from pe is no longer necessary, filter retrieval should be considered.Filter retrieval should be attempted when feasible and clinically indicated.Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e.G., a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications).For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set.It may also be retrieved with the cloversnare® vascular retriever.Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques.The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position.The safety or effectiveness of these alternative retrieval techniques has not been established.Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve.For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.Unknown if the reported anxiety, depression, bipolar, worry, fear, physical limitations is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key12452810
MDR Text Key270792769
Report Number1820334-2021-02131
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexFemale
Patient Weight75 KG
-
-