| Model Number |
383400 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation. if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device's flow was not conforming.Additional information was provided stating that the issue happened during reconditioning test.The device flowed faster than the set rate.
|
| |
|
Additional Manufacturer Narrative
|
|
The service history record was reviewed and indicated that this is the first time that this kangaroo joey enteral feeding pump has been returned to the service center.The pump was decontaminated with an isopropyl alcohol wipe and an evaluation of the pump was performed.The unit was tested, and the reported issue could not be confirmed at this time.However, the software, battery, and overlay were found to be old and were replaced for this reason.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be used for tracking and trending purposes.
|
| |
|