• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES BRONCHOFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES BRONCHOFIBERSCOPE Back to Search Results
Model Number BF-N20
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code Fungal Infection (2419)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device referenced in this report has not yet been received by olympus for physical evaluation (although it is anticipated).The definitive cause of the user¿s experience cannot be determined at this time.The investigation is ongoing.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.This report has been submitted by the importer on mdr# 2951238-2021-00401.
 
Event or Problem Description
It is reported by the customer, an unspecified time after a procedure using one of their 10 bronchoscopes, preliminary findings of fungal culture specimen collected from the patient via bronchoalveolar lavage (bal) from right middle lobe (rml) of lung shows: scant growth gram positive branching rods suggestive of aerobic actinomycete (originally reported as mold).It is unknown what treatment the patient required as a result.The patient¿s current condition is unknown.The customer states that it is not known which of their 10 bronchoscopes was used in the case, as the surgery staff did not document this information in the procedure record.For this reason, the customer is requesting all 10 bronchoscopes be evaluated to determine if any of them are potentially a contributing factor to the positive patient culture.Additional details regarding the patient and event have been requested.At this time, no further details have been provided.Case with patient identifier (b)(6) reports a bf-xp60.Case with patient identifier (b)(6) reports a bf-3c40.Case with patient identifier (b)(6) reports a bf-3c40.Case with patient identifier (b)(6) reports a bf-3c40.Case with patient identifier (b)(6) reports a bf-3c40.Case with patient identifier (b)(6) reports a bf-n20.Case with patient identifier (b)(6) reports a bf-3c40.Case with patient identifier (b)(6) reports a bf-n20.Case with patient identifier (b)(6) reports a bf-3c40.Case with patient identifier (b)(6) reports a bf-xp60.The customer further reported they had been sending their scopes to (b)(4) for repair.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause cannot be identified.This device was not returned or evaluated, though the device for patient identifier (b)(6) was evaluated where it was found that the bending section cover and glue was non-olympus, there was restriction in the forceps and brush passage due to a kink, the image focus was off, the camera switch was not working, the bending section and insertion tube were dented and non-olympus, the light guide tube was non-olympus, and the eyepiece body was discolored.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES BRONCHOFIBERSCOPE
Common Device Name
BRONCHOFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12455473
Report Number8010047-2021-11573
Device Sequence Number17347089
Product Code EOQ
UDI-Device Identifier04953170051142
UDI-Public04953170051142
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K910423
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-N20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2021
Supplement Date Received by Manufacturer02/03/2022
Initial Report FDA Received Date09/10/2021
Supplement Report FDA Received Date02/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/17/1996
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age15 MO
Patient SexMale
-
-