Model Number 154004A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the urine leaked from the shaft between the tamper evident seal and the tip on the 6th day of use.
 
Event or Problem Description
It was reported that the urine leaked from the shaft between the tamper evident seal and the tip on the 6th day of use.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for treatment.The product caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), 1 two-way silicone foley catheter.Visual inspection of the sample noted obvious visible defects.The foley catheter shaft had a hole on the return sample.A potential root cause for this failure could be "tooling damaged (core pins)".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[warnings] 1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients ·patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (4) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients patients with known allergy to silver coated catheter." h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12455568
Report Number1018233-2021-05579
Device Sequence Number4081888
Product Code EZL
UDI-Device Identifier00801741029554
UDI-Public(01)00801741029554
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2024
Device Model Number154004A
Device Catalogue Number1758SI16
Device Lot NumberMYEZ3719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/19/2021
Supplement Date Received by Manufacturer02/04/2022
Initial Report FDA Received Date09/10/2021
Supplement Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-