Manufacturer Narrative
|
Occupation: non-healthcare professional investigation: the reported allegations have been investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Catalog number and lot number are unknown.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
|
Event Description
|
The following information is alleged: the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2015.The patient underwent a computed tomography (ct) approximately two years and five months later, which revealed the filter had moved and several struts perforated through the ivc wall.Hospital and medical records have been requested, but not yet provided.
|
|
Manufacturer Narrative
|
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2015 via the right internal jugular vein due to deep vein thrombosis (dvt) and pulmonary embolism (pe).Patient is alleging vena cava perforation.Report from ct (computed tomography): "inferior vena cava filter in place.The prongs extend beyond the inferior vena cava in the retro peritoneal fat.No associated retroperitoneal hemorrhage.".
|
|
Manufacturer Narrative
|
Investigation: investigation is reopened due to additional information provided.The following allegations have been investigated: organ/aorta perforation, tilt.The reported allegations have been further investigated based on the information provided to date.The additional information regarding organ/aorta perforation does not change the previous investigation results for vena cava perforation.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.(b)(4) devices in lot.One other complaint has been reported against the lot for a different issue.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Per ct report, "filter position: below above the renal veins." "filter tilt: yes." "filter penetration: yes." "impressions: the filter is tilted posterior with its proximal cone against the ivc wall.Axial image 34.The anterior strut penetrates 10 mm through the ivc wall.It penetrates into the duodenum.Sagittal image 45.The left strut penetrates 12 mm through the ivc wall.It penetrates into the aorta.Coronal image 45.The posterior strut penetrates 8 mm through the ivc wall.Sagittal image 43.The right strut penetrates 12 mm through the ivc wall.Coronal image 42.Two arms each penetrate 8 mm through the ivc wall.".
|
|