H10: four (4) devices were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed as follows, remove protector, ensure slide clamp is open, prime set, purge air, attach male luer until secure, open and close slide clamp as required.Additionally, clear passage and pressure tests were performed on the samples.As a result, the clear passage test failed in three (3) units and the pressure test could not be performed in the three units due to the clear passage test failure.The reported condition was not verified on one of the samples however, the reported condition of no flow was verified on the other three (3) samples.The cause of the condition was determined to be a machine related issue during the manufacturing process.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|