ZEPHYR ENDOBRONCHIAL DELIVERY CATHETER; ENDOBRONCHIAL VALVE Back to Search Results
Model Number ZEPHYR 4.0 EDC
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2021, the patient underwent a bronchoscopic lung volume reduction (blvr) procedure with zephyr valves.The physician had used the zephyr 4.0 endobronchial delivery catheter to size airways in the right upper lobe (rul) and right middle lobe (rml) and deliver three valves.One of the sizing wings of the catheter was detached, and the fragment was not found.It was not clear if the sizing wing fell off inside the patient, was inside of the bronchoscope and subsequently removed, or if it was removed from the patient while mucus and blood were cleared from the airway.The physician had difficulty accessing the apical segment of the rul.The physician used a new zephyr 4.0-j catheter and was able to successfully complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR ENDOBRONCHIAL DELIVERY CATHETER
Common Device Name
ENDOBRONCHIAL VALVE
MDR Report Key12457153
Report Number3007797756-2021-00244
Device Sequence Number17160475
Product Code NJK
UDI-Device Identifier00811907030522
UDI-Public(01)00811907030522(10)505292V50(17)221116
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P180002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2022
Device Model NumberZEPHYR 4.0 EDC
Device Catalogue NumberEDC-TS-4.0
Device Lot Number505292-V5.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2021
Initial Report FDA Received Date09/10/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-