| Model Number |
FDS50020 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during the procedure that the subject stent prematurely deployed while attempting to re-sheath.The physician replaced it with a new device and completed the procedure without clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and the lot number was confirmed with the packaging returned with the device.During visual inspection the device was returned together with catheter.The stent delivery wire was found to be kinked/bent.The deployed stent was found to be inside returned catheter.The stent was removed using patency mandrel.The stent wire ends were found to be damaged.The introducer sheath was not returned.The returned catheter was intact.Functional testing was not performed.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event was confirmed during the device analysis.The device failed to meet specification when returned for analysis.It was reported that the stent deployed half was outside the catheter and half was inside.It is probable that the device was damaged during the clinical procedure, causing the reported event.An assignable cause of procedural factors will be assigned to the reported event of the stent deployed prematurely during use and to the analyzed event of the stent delivery wire kinked/bent, the stent deformed and the stent deployed prematurely during use, as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural and/or anatomical factors during use.
|
| |
|
Event or Problem Description
|
|
It was reported during the procedure that the subject stent prematurely deployed while attempting to re-sheath.The physician replaced it with a new device and completed the procedure without clinical consequences to the patient.
|
| |
|
Search Alerts/Recalls
|