Model Number IPN915189 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device investigation is pending.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
Loose screw on jaw area, shaft bent and unable to remove trocar.
|
|
Manufacturer Narrative
|
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet , inc.Kenosha wi facility as part of a (b)(4) pc.Lot in june of 2017.Evaluation of the returned instrument shows that the tips are loose and misaligned and tilted to one side and the jaw pivot pin is pushed thru one side of the damaged/bent outer tube assembly.There is a surgical trocar still installed on the outer tube assembly.We are able to validate this complaint.We are unable to determine what caused the jaws to become loose and misaligned and for the outer tube assembly to be damaged/bent and for the jaw pivot pin to be pulled thru one side of the outer tube assembly but mishandling of this device at the end user's facility is suspected.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings , no further actions will be taken in response to this complaint and this record will be deemed closed.
|
|
Event Description
|
Loose screw on jaw area, shaft bent and unable to remove trocar.
|
|
Search Alerts/Recalls
|
|
|