| Model Number |
L20310 |
| Medical Device Problem Code |
Expiration Date Error (2528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the kho size 10 corail stem was found to have a package expiration date of 6/30/2021 after patient was closed and in recovery.Doe: (b)(6) 2021 affected side: right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4) investigation summary = > no device associated with this report was received for examination.No distribution error identified.Root cause: user error for not having verified sterile expiry date before having used the device.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot = > the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|