| Model Number |
21-4485-24 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluation- the device was returned for evaluation.The device was examined; this showed the catheter was still attached to the portal; however a crack was found along the length of the catheter.The device was given functional testing and no leakage was found.While a crack was found in the material no functional issues were identified.The manufacturing process includes a patency test for the catheter to verify no damage prior to packaging.The reported issue could not be attributed to a manufacturing issue; the likely cause was determined to have occurred due to stress during use.
|
| |
|
Event or Problem Description
|
|
It was reported the device was in patient use.The device was removed due to a hole found in catheter that was identified upon fluoroscopy.The surgical removal occurred with no further adverse effects reported.
|
| |
|