• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION BAND 620BG25 DURAN ANCORE ANNULOPLASTY C; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION BAND 620BG25 DURAN ANCORE ANNULOPLASTY C; RING, ANNULOPLASTY Back to Search Results
Model Number 620BG25
Medical Device Problem Codes Gradient Increase (1270); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Mitral Valve Insufficiency/ Regurgitation (4451); Insufficient Information (4580)
Date of Event 08/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during the implant of this mitral annuloplasty band, it was explanted and replaced with a non-medtronic valve.It was reported that the repair attempt with the annuloplasty device was not successful due to the patient's anatomy, and that the event was not related to the product.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Updated patients age.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic received additional information that post implant of this mitral annuloplasty band, there was mild regurgitation still present.The patient was removed from bypass, where mild reguritation remained, and there was a gradient of 5mmhg across the valve.The physican was going to leave this as is, however the patient later developed increased mild to severe regurgitation and blunting of the pulmonary vein flow.As a result, the surgeon explanted the band and replaced it with a non medtronic bioprosthetic valve which alleviated the regurgitation and lowered the gradients.No additional adverse patient effects were reported.  medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAND 620BG25 DURAN ANCORE ANNULOPLASTY C
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key12461450
Report Number2025587-2021-02801
Device Sequence Number4067151
Product Code KRH
UDI-Device Identifier00643169793446
UDI-Public00643169793446
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K032810
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/25/2026
Device Model Number620BG25
Device Catalogue Number620BG25
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/17/2021
Supplement Date Received by Manufacturer09/15/2021
09/15/2021
Initial Report FDA Received Date09/13/2021
Supplement Report FDA Received Date09/22/2021
09/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient Weight123
-
-