Medtronic received information reporting that on (b)(6) 2021, a pipeline flex with shield (model: ped2-450-20, lot: b209283) was being implanted in attempts to treat 2 unruptured supraclinoid left internal carotid artery (ica) aneurysms.After the pipeline was delivered, there was some foreshortening resulting in incomplete neck coverage of the more distal aneurysm.There was no friction or difficulty during delivery or positioning.The tip of the catheter was not moved during deployment.The pipeline did not jump during deployment but missed the landing zone.A second, shorter pipeline flex with shield (model: ped2-450-12, lot: b208340) was then placed distally in the proximal m1 for full coverage.Both devices opened well with good apposition seen in final angio.The patient was prescript dual antiplatelet treatment (dapt) - aspirin and brillinta.2 days after discharge, the patient began experiencing stroke symptoms and was readmitted on (b)(6) 2021.On (b)(6) 2021, the patient was admitted to intensive care (icu) and a thrombectomy was done.Occlusion recurred on (b)(6) 2021 and another thrombectomy was successfully performed.
|