• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. 6.5 X 50MM MULTIAXIAL SCREW; EBI 5.5 HELICAL FLANGE SPINAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. 6.5 X 50MM MULTIAXIAL SCREW; EBI 5.5 HELICAL FLANGE SPINAL SYSTEM Back to Search Results
Catalog Number 2000-2450
Medical Device Problem Codes Material Twisted/Bent (2981); Positioning Problem (3009)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported a revision was performed to remove a pedicle screw because a scan showed the screw had breached the pedicle wall.When the screw was removed, the shaft of the screw was very slightly bent.A new screw was placed during the revision.
 
Event or Problem Description
It was reported a revision was performed to remove a pedicle screw because a scan showed the screw had breached the pedicle wall.When the screw was removed, the shaft of the screw was very slightly bent.A new screw was placed during the revision.
 
Additional Manufacturer Narrative
Device evaluation visual inspection revealed the screw shank is bent.The tulip head is also stuck in place.Potential cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to an off-axis force placed on the screw during the original surgery or the screw not being sized/positioned correctly during the original surgery.Dhr review per dhr review, the part was likely conforming when it left zimmer biomet control.Device use this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5 X 50MM MULTIAXIAL SCREW
Common Device Name
EBI 5.5 HELICAL FLANGE SPINAL SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key12463837
Report Number3012447612-2021-00346
Device Sequence Number4815184
Product Code MNH
UDI-Device Identifier00888480207764
UDI-Public(01)00888480207764(10)J6783199
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K061441
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2000-2450
Device Lot NumberJ6783199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2021
Supplement Date Received by Manufacturer11/08/2021
Initial Report FDA Received Date09/13/2021
Supplement Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-