• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN TRUSEAL ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN TRUSEAL ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number M635TU70010
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Tamponade (2226); Cardiac Perforation (2513); Pericardial Effusion (3271)
Date of Event 08/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a pericardial effusion occurred.A left atrial appendage (laa) closure procedure was being performed.No pericardial effusions were noted on transesophageal echocardiogram at the start of the procedure.The 24mm watchman flx delivery system was not yet to the proximal marker band of the watchman truseal access system (was) when a perforation was noted.A pericardial effusion with cardiac tamponade occurred.A contrast injection was performed showing the was advanced distally in the laa.A pericardial tap was performed followed by emergent surgical intervention.The emergent surgery was successful and then the patient was transferred to another hospital for post cardiac care since the implanting hospital did not have it.The patient was doing well from the complications and surgical repair, but had several other unknown issues going on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN TRUSEAL ACCESS SYSTEM
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12464659
Report Number2134265-2021-10894
Device Sequence Number16455104
Product Code DQY
UDI-Device Identifier08714729965701
UDI-Public08714729965701
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/22/2022
Device Model NumberM635TU70010
Device Catalogue NumberM635TU70010
Device Lot Number0023683944
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/20/2021
Initial Report FDA Received Date09/13/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
-
-