This complaint was reopened in reference to (b)(4).Manufacturing site evaluation: manufacturing and quality control data.The progav® was manufactured by a qualified employee in september 2013.Deviations during assembly did not occur.The valve was sterilized by miethke and released for shipment after final inspection.The progav® has a normal pressure range of 0 to 20 cmh2o.The valve was inspected as article fv425t.All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specification.Investigative statement: on the basis of juristic requirements, the surgeon must provide a personal request as a step of safeguard.Unfortunately, no consent was received from the surgeon.An official release is missing for the investigation of the product.For this reason, the product was returned for discharge without examination.Result: as described in scientific literature, the problem encountered is one of the known, inevitable risks of hydrocephalus (hc)-therapy by shunt implants.Manufacturing defects at the time of release have been excluded.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required.
|