• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2N1333
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event, concomitant medical products: this event occurred during the unspecified date of (b)(6) 2021.Initial reporter name and address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an unspecified quantity of non-dehp standard bore catheter extension sets were ¿block/clogged¿ and would not flow.It was further reported that it was impossible to empty the sets unless by injecting liquid with an unspecified syringe by applying strong pressure on the plunger.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was not received for evaluation; therefore, a device analysis could not be completed.However, the condition of no flow, as reported in the event description, was determined to be a machine related issue during the manufacturing process.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
MDR Report Key12465708
Report Number1416980-2021-05648
Device Sequence Number4891806
Product Code FPA
UDI-Device Identifier00085412475394
UDI-Public(01)00085412475394
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
K132734
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number2N1333
Device Lot NumberUR19K17076
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/18/2021
Supplement Date Received by Manufacturer10/07/2021
Initial Report FDA Received Date09/13/2021
Supplement Report FDA Received Date10/13/2021
Date Device Manufactured11/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNSPECIFIED SYRINGE; UNSPECIFIED SYRINGE
-
-