If the information is unknown, not available or does not apply, the section/field of the form is left blank.The depth gauge for 2.0mm and 2.4mm screws (p/n: 319.006, lot number: 6312972) was received at us cq.Upon visual inspection, it was observed the needle component of the device had broken off and was not returned.Also, the protection sleeve was not returned with the complaint device.Rest of the surface of the device shows normal wear consistent with the device use which would not contribute to the complaint condition.This complaint is confirmed as the needle component of the complaint device had broken off.No definitive root cause could be determined based on the provided information.The reported condition of fell apart cannot be confirmed.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part # 319.006, synthes lot # 6312972, supplier lot # na, release to warehouse date: 03 feb 2010, manufactured by (b)(4), no ncr's were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
It was reported that on an unknown date, during inspection, the graphic case, t-handle and depth gauge became detached.There was no patient involvement.This report is for one (1) depth gauge for 2.0mm and 2.4mm screws.This is report 1 of 1 for (b)(4).
|