Model Number 319.006
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.The depth gauge for 2.0mm and 2.4mm screws (p/n: 319.006, lot number: 6312972) was received at us cq.Upon visual inspection, it was observed the needle component of the device had broken off and was not returned.Also, the protection sleeve was not returned with the complaint device.Rest of the surface of the device shows normal wear consistent with the device use which would not contribute to the complaint condition.This complaint is confirmed as the needle component of the complaint device had broken off.No definitive root cause could be determined based on the provided information.The reported condition of fell apart cannot be confirmed.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part # 319.006, synthes lot # 6312972, supplier lot # na, release to warehouse date: 03 feb 2010, manufactured by (b)(4), no ncr's were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, during inspection, the graphic case, t-handle and depth gauge became detached.There was no patient involvement.This report is for one (1) depth gauge for 2.0mm and 2.4mm screws.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key12467777
MDR Text Key271275354
Report Number2939274-2021-05408
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982189943
UDI-Public(01)10886982189943
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number319.006
Device Catalogue Number319.006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/09/2021
Initial Date FDA Received09/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
GRAPHIC CASE/2/3 LNTH 2 BAY/F/MOD GC SYS; T-HANDLE WITH QUICK COUPLING
-
-