| Model Number |
OWREN'S VERONAL BUFFER |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer contacted a siemens customer care center (ccc).The glass of the bottle was broken structurally, caused by a major impact.The label was applied correctly, and the glass pieces were attached to the label as well, indicating that the vial was not damaged during labeling.There is no indication of a systematic failure of processes or any material issues.The most likely cause of the event was a shipping and handling issue.The cause of the event is unknown.The reagent is performing according to specifications.No further evaluation of this device is required.
|
| |
|
Event or Problem Description
|
|
While opening a bottle of owren's veronal buffer (ovb), the customer received a cut on their finger after the bottle shattered in their hand.The customer cleaned the wound and did not seek medical attention.The box from which the broken bottle was taken had wet marks on it and looked physically damaged.The customer discarded the remaining bottles of ovb in the box.There are no known reports of adverse health consequences due to this event.
|
| |
|