• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT4
Medical Device Problem Code Ejection Problem (4009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens and the qualified company cartridge were returned inside the opened lens carton along with the disassembled lens case.Insufficient viscoelastic was observed in the cartridge.Stress lines were observed at the tip.The cartridge has evidence of being placed into a handpiece.The lens was returned in a small clear bag.Small bits of solution was dried on the lens.No damage was observed to the lens.The lens was cleaned for further evaluation.A dimensional inspection (plan view) was conducted.The lens was within the specification per the approved template.The cartridge was cleaned for further evaluation.Top coat dye stain test was conducted with acceptable results.Product history records were reviewed and documentation indicated the product met release criteria.A non-qualified viscoelastic was indicated.The root cause of the reported complaint may be related to a failure to follow the ifu.A non-qualified viscoelastic was indicated.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Company recommends using the qualified company iol delivery system or any other company qualified combination.An inadequate amount of viscoelastic was observed in the cartridge.The ifu also instructs to completely fill the cartridge with ovd(ophthalmic viscosurgical devices) (diagram provided) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Top coat dye stain test was conducted with acceptable results.` the manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse unit manager reported that the lens was early out of the cartridge at the level of the main incision with patient contact.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key12470755
Report Number1119421-2021-01748
Device Sequence Number11710880
Product Code MJP
UDI-Device Identifier00380652250498
UDI-Public00380652250498
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/19/2025
Device Model NumberSN6AT4
Device Catalogue NumberSN6AT4.235
Device Lot Number15116910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/24/2021
Initial Report FDA Received Date09/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; MONARCH IIIC CARTRIDGE; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE; PROVISC OVD
-
-