SMITH & NEPHEW, INC. GII MOD PAT REAMER GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
71440311 |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned gii mod pat reamer gde confirms the pin that holds the device together in the middle is broken.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that the supporting arm of a genesis ii modified patellar reamer guide is broken.In addition, it was noted that the pin fell out.As this was noticed in a non-surgical environment, there was no patient involvement.
|
| |
|
Search Alerts/Recalls
|
|
|