| Model Number |
392412 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device is at early eri indication with unexpected battery behavior.No adverse patient events have been reported.The device will be changed out.Should additional information be received, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.Please note this icd is affected by a field safety corrective action, bio-lqc, initiated in march 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was explanted (b)(6) 2021.We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.Please note this icd is affected by a field safety corrective action, bio-lqc, initiated in march 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was explanted (b)(6) 2021.Upon receipt, the device was interrogated.The interrogation could be properly performed and revealed the mos1 battery status.Battery trend data showed a temporary voltage drop followed by recovery of the battery voltage.An unexpected battery voltage trend could be confirmed during analysis.Please note, that this icd is subjected to the field safety corrective action, bio-lqc, initiated in march 2021.
|
| |
|