BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.5/35; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 391235
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Unexpected Color (4055)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment.During unpacking it was noticed that the stent was detached from the delivery system.It was also mentioned that the color was not as usual.
 
Additional Manufacturer Narrative
Combination product: yes.The complaint instrument was not returned for analysis.Therefore, no technical investigation on the subject could be performed.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the product release documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent as well as surface and coating defects.The different colors observed on orsiro stents are caused by the thickness variation of the passive probio coating.Typically the color of the stents varies between bluish, purple and yellowish.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.5/35
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key12474414
Report Number1028232-2021-05123
Device Sequence Number3392519
Product Code NIQ
UDI-Device Identifier07640130439368
UDI-Public07640130439368
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/07/2023
Device Model Number391235
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03210598
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/14/2021
Supplement Date Received by Manufacturer09/22/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date09/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-