Model Number PAPS3
Medical Device Problem Codes Crack (1135); Suction Problem (2170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure using a penumbra engine canister (canister) and a penumbra engine (engine).During the procedure, one of the engine indicator lights was not illuminated.The canister and aspiration tubing (tubing) were disconnected and re-connected twice but the issue did not resolve.It was reported that the canister lid felt loose.Therefore, the canister was removed.After the canister was returned to the distributor, it was found that the lid was cracked.The procedure was completed using a new canister and the same engine.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned canister could not confirm the reported complaint.The lid was inspected for a crack, but the observed damaged was determined to be scuff marks.The cause of this damage could not be determined, and the damage did not affect the canister''s ability to hold vacuum pressure.During functional testing, the canister was tested on a demonstration engine.The engine was able to achieve vacuum pressure within specification, and all four vacuum indicator lights illuminated.The root cause of the reported complaint could not be determined.The supplier performs 100% visual and functional inspection prior to shipping to penumbra.Penumbra performs packaging and label inspection upon receipt prior to shipment to customer.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA ENGINE CANISTER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12474559
Report Number3005168196-2021-02036
Device Sequence Number4892640
Product Code NRY
UDI-Device Identifier00814548019758
UDI-Public00814548019758
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K180008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPAPS3
Device Catalogue NumberPAPS3
Device Lot NumberS11880
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/25/2021
Supplement Date Received by Manufacturer10/07/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date10/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age37 YR
Date Report Sent to FDA01/01/2005
-
-