Catalog Number 3910-600-062
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.All pieces were retrieved.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor broken/ probable root cause: design/ insufficient material strength/ anchor assembly design not strong enough to withstand impaction/ anchor geometry too blunt for effective bone penetration/ process: anchor not manufactured to specification.Improper assembly of anchor onto inserter.Application: hard bone.Excessive force.Incorrect angle of attack.No pilot hole prepared in the event of hard bone.Incorrect instrumentation used to prepare pilot hole.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.All pieces were retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REELX STT 4.5MM KNOTLESS ANCHOR
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key12474771
Report Number0002936485-2021-00504
Device Sequence Number14266692
Product Code MBI
UDI-Device Identifier37613252592809
UDI-Public37613252592809
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K120824
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-600-062
Device Lot Number21042AG2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/17/2021
Supplement Date Received by Manufacturer08/17/2021
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-