| Catalog Number |
3910-600-062 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor broken.Probable root cause: design: insufficient material strength.Anchor assembly design not strong enough to withstand impaction.Anchor geometry too blunt for effective bone penetration.Process: anchor not manufactured to specification.Improper assembly of anchor onto inserter.Application: hard bone.Excessive force.Incorrect angle of attack.No pilot hole prepared in the event of hard bone.Incorrect instrumentation used to prepare pilot hole.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure.All pieces were retrieved.
|
| |
|
Search Alerts/Recalls
|