BECTON DICKINSON TRAY SPN WHIT25G3.5 L/B-D/E BLUE DRAPE; ANESTHESIA CONDUCTION KIT Back to Search Results
Model Number 405672
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Oem manufacture: (b)(4).Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the tray spn whit25g3.5 l/b-d/e blue drape was ineffective, and the treatment failed to work.The following information was provided by the initial reporter: there were several consecutive failed spinal anesthetics on ob.Consistent factor has been using the spinal bupivacaine (0.75%) from the kits.
 
Event or Problem Description
It was reported that the tray spn whit25g3.5 l/b-d/e blue drape was ineffective, and the treatment failed to work.The following information was provided by the initial reporter: there were several consecutive failed spinal anesthetics on ob.Consistent factor has been using the spinal bupivacaine (0.75%) from the kits.
 
Additional Manufacturer Narrative
H6: investigation summary: one sample was provided to our quality team for investigation.The anesthetic ampules and contents were examined.No discernible haze or particulate matter was observed.The samples appeared clear, colorless and did not contain any coloration or hue and had the appearance of water.The returned samples met the criteria for passing, therefore the reported failure was not confirmed.A review of the internal manufacturing device records and raw material history files for the reported lot number 0001387936 was performed and no recorded quality problems or rejections related to this incident were found.The sterilization records indicated no issues.Product undergoes inspections during manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.Based on the available information we are not able to identify a root cause at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAY SPN WHIT25G3.5 L/B-D/E BLUE DRAPE
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12475278
Report Number2243072-2021-02320
Device Sequence Number4739471
Product Code CAZ
UDI-Device Identifier00382904056728
UDI-Public00382904056728
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
DISCRETION
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2021
Device Model Number405672
Device Catalogue Number405672
Device Lot Number0001387936
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2021
Supplement Date Received by Manufacturer02/18/2022
Initial Report FDA Received Date09/15/2021
Supplement Report FDA Received Date02/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-